3He proportional counter in polyethylene moderator
Energy range
PU2 (γ)
60 keV ... 3 MeV
BDKN-03 (n)
0.025 eV ... 14 MeV
Measurement range of ambient radiation dose rate equivalent
PU2 (γ)
1 μSv/h ... 10 mSv/h
BDKN-03 (n)
0.1 μSv/h ... 10 mSv/h
Measurement range of ambient radiation dose equivalent
PU2 (γ)
1 μSv ... 1 Sv
BDKN-03 (n)
0.1 μSv ... 10 Sv
Energy dependence relative to 662 keV (137Cs) [PU2 (γ)]
-25%...+35%
(for energy range from 60 keV to 3 MeV)
Sensitivity to137Cs gamma radiation [PU2]
1.0 cps/μSv?h-1
Sensitivity to Pu-Be neutron radiation [BDKN-03]
0.355 cps/μSv?h-1
Neutron flux density measurement range [BDKN-03]
0,1 … 104neutron·s-1·cm-2
Neutron fluence measurement range [BDKN-03]
1 … 3·106neutron·cm-2
Sensitivity to Pu-Be neutron radiation [BDKN-03]
1.5 cps/(neutron·s-1·cm-2)
Limit of intrinsic relative measurement error
±20%
Continuous run time
≥24 h
Working temperature range
-30°С ... +50°С
Protection class
IР64
Power supply
Integrated rechargeable battery pack;
+12V external power supply unit;
220 V 50 Hz external power supply unit
Overall dimensions, weight
PU2 (γ)
200х85х36 mm, 0.5 kg
BDKN-03 (n)
314х220х263 mm, 7.8 kg
Relative sensitivity coefficients for standard neutron radiation sources with different energiesfor ambient dose equivalent rate measurement
n-radiation source
BDKN-03
Thermal,
En=0.025 eV
0,225±0,045
Ra-Be,
En=100 keV
0,81±0,08
252Cf,
En=2.13 MeV
1,02±0,10
Pu-Be,
En=4.16 MeV
1,0
Relative sensitivity coefficients for standard neutron radiation sources with different energiesfor flux density measurement
n-radiation source
BDKN-03
Thermal,
En=0.025 eV
0,0064±0,0013
Ra-Be,
En=100 keV
0,182±0,018
252Cf,
En=2.13 MeV
1,01±0,10
Pu-Be,
En=4.16 MeV
1,0
Radiation monitor meets requirements of the following International standards: IEC 60846-1:2009, IEC 60846-2:2007,IEC 60325:2002, IEC 61005:2003, and the following safety standards: IEC 61010-1:1990, and Electromagnetic compatibility requirements of: EN 55022:1998+A1:2000+A2:2003, EN 55024:1998+A1:2001+A2:2003