Specification |
AT6104DM |
AT6104DM1 |
Gamma radiation detector |
NaI(Tl) scintillator, Ø63x63 mm |
NaI(Tl) scintillator, Ø63x160 mm |
Registered gamma radiation energy range |
50keV ... 3MeV |
Identified radionuclides |
137Cs, 134Cs, 131I, 40K, 226Ra, 232Th, 60Co, 24Na, 54Mn, 65Zn и др. |
Measurement range of specific activity in water for 4π measurement geometry: |
|
|
134Cs, 137Cs, 131I, 60Co |
3 ... 1·106 Bq/kg |
1 ... 1·106 Bq/kg |
40K |
250 ... 2·104 Bq/kg |
100 ... 2·104 Bq/kg |
Measurement range of bottom sediments specific activity for 2π measurement geometry: |
|
|
134Cs, 137Cs, |
50 ... 1·105 Bq/kg |
20 ... 1·105 Bq/kg |
Intrinsic relative error of activity measurement |
±30% max. |
Measurement range of ambient radiation dose rate equivalent |
0.01 ... 100 μSv/h |
Intrinsic relative error of ambient dose equivalent rate measurement |
±20% max. |
Energy dependence relative to 662 keV (137Cs) |
±20% max. |
Anisotropy in angular spacing ±120° relative to vertical axis (100 keV ... 3 MeV energy range) |
±30% |
Sensitivity to gamma radiation |
|
|
241Am |
4050 cps/μSv?h-1 |
11800 cps/μSv?h-1 |
137Сs |
2350 cps/μSv?h-1 |
5100 cps/μSv?h-1 |
60Сo |
1300 cps/μSv?h-1 |
2900 cps/μSv?h-1 |
Typical resolution at 662 keV (137Cs) |
8% |
9% |
Maximum input statistical load |
≥5·104 s-1 |
Number of ADC channels |
512 or1024 |
512 |
Operation mode setup time |
2 min |
Integral nonlinearity |
≤1% |
Continuous run time in normal conditions |
≥9 h |
Measurement instability during continuous service |
≤5% |
Working temperature range |
From -20°C to +50°C |
Protection class of the detection unit |
IР67 (Withstands static hydraulic pressure up to 5 MPa) |
GPS |
GPS receiver is integrated into PC. Positioning accuracy ≥3 m |
Connection to PC |
RS 422 |
Overall dimensions and weight of the detection unit |
Ø130x510 mm, 4.5 kg |
Ø130x633 mm, 6.5 kg |