Detectors |
|
Gamma channel |
NaI(Tl) ø63х63 mm |
Beta channel |
Plastic scintillator, ø128x9 mm |
Energy range |
|
Gamma radiation |
50 keV...3 MeV |
Beta radiation |
150 keV...3.5 MeV |
Volumetric (specific) activity measuring range without sample concentration (in Spectrometric and Radiometric modes) |
|
137Cs |
2...106 Bq/l (Bq/kg) |
40K |
20...2·104 Bq/l (Bq/kg) |
226Ra |
3...104 Bq/l (Bq/kg) |
232Th |
3...104 Bq/l (Bq/kg) |
90Sr (In Radiometric mode only) |
20...106 Bq/l (Bq/kg) |
131I (In Spectrometric mode only) |
10...105 Bq/l (Bq/kg) |
134Cs (In Spectrometric mode only) |
6...105 Bq/l (Bq/kg) |
Intrinsic relative error of activity measurement with Р = 0.95 |
±20% max. |
Measured sample density range |
0.2...1.6 g/cm3 |
Lower limit of 90Sr measurement range with sample concentration in conversion to "wet" sample |
|
For potable water |
0.2 Bq/l |
For milk, baby food |
1.5 Bq/l |
For potatoes, corn, grain, agricultural raw materials |
2.0 Bq/kg |
Integral nonlinearity |
<1% |
Relative energy resolution for 662 keV gamma-ray line |
≤7.5% |
Maximum input statistical load |
1·104 s-1 |
Calibration scale instability during continuous service |
<2% |
Measurement instability during continuous service |
<5% |
Number of ADC channels |
1024 |
Continuous run time |
≥24 h |
Operation mode setup time |
<15 min |
Working temperature range |
+10°С...+35°С |
Relative humidity with air temperature ≤30°C without condensation |
≤75% |
Power supply |
230 VAC (±23 V), 50±0.4 Hz |
Power consumption without printer |
300 VA max. |
Overall dimensions, weight |
|
Gamma detection unit |
ø98x371 mm, 3 kg |
Beta detection unit |
ø138x340 mm, 3 kg |
Protection unit |
ø600x950 mm, 120 kg |
Measurement vessels volume |
|
For "wet" samples |
Marinelli beaker, 1 l; Flat vessel, 0.5 and 0.1 l |
For concentrated samples |
Flat vessel, 0.2 l and 0.03 l |